Attenuation Coefficient and Mean Free Path of 60, 290, 404, 566 and 662 keV -rays for PbWO4 and CdWO4 Scintillation Crystals

Authors

  • Krittiya Sreebunpeng Department of Physics, Faculty of Science, Chandrakasem Rajabhat University
  • Weerapong Chewpraditkul Department of Physics, King Mongkut’s University of Technology Thonburi
  • Nimit Kimpraphan Department of Physics, Faculty of Science, Chandrakasem Rajabhat University
  • Parnuwat Chimalawong Department of Physics, Faculty of Science, Chandrakasem Rajabhat University
  • Korravee Wongwigkarn Department of Physics, Faculty of Science, Chandrakasem Rajabhat University

Keywords:

Compton scattering, Attenuation coefficient, CdWO4, PbWO4

Abstract

In this work, the attenuation coefficient and mean free path of 60, 290, 404, 566 and 662 keV gif.latex?\gamma-rays for PbWO4 and CdWO4 scintillation crystals were measured using the transmission method. The 60 and 662 keV gif.latex?\gamma-rays were obtained from 241Am and 137Cs radioactive sources, respectively. The 290, 404 and 566 keV g-rays were obtained from the Compton scattering of 662 keV gif.latex?\gamma-rays (137Cs source) at scattering angle of 90°, 60° and 30°, respectively. For comparison, the attenuation coefficient and mean free path of gamma rays for the studied crystals were also calculated over energy range 1 – 1000 keV using the WinXCom program.

References

Nikl, M. (2006). Scintillation detectors for x-rays. Measurement Science and Technology, 17(4), R37-R54.

van Eijk and C.W. (2002). Inorganic scintillators in medical imaging. Physics in Medicine & Biology, 47(8), R85-106.

Chen, S., Bonurham, M., Rabiei A. (2014). Novel light-weight materials for shielding gamma ray. Radiation Physics and Chemistry, 96, 27-36.

Lecoq, P., Dafinei, I., Auffray, E., Schneegans, M., Korzhik, M.V. Missevitch, O.V, Pavlenko, V.B., Fedorov, A.A., Annenkov, A.N., Kostylev, V.L., Ligun, V.D. (1995). Lead tungstate (PbWO4) scintillators for LHC EM calorimetry. Nuclear Instruments and Methods in Physics Research A, 365, 291-298.

Liu, X., Hu, G., Feng, X.M., Huang, Y., Zhang Y. (2002). Influence of PbF2 Doping on Scintillation Properties of PbWO4 Single Crystals. Physics status Solidi (a), 190, R1-R3.

Lecoq, P. (2016). Development of new scintillators for medical applications. Nuclear Instruments and Methods in Physics Research A, 809, 130-139.

Holl, I., Lorenz, E., Mageras, G. (1988). A Measurement of the Light Yield of Common Inorganic Scintillators. IEEE Transactions on Nuclear Science, 35, 105-109.

Melcher, C.L. (2005). Perspective in the future development of new scintillators. Nuclear Instruments and Methods in Physics Research A, 537, 6-14.

Biswas, R., Sahadath, H., Mollah, A.S., Huq, F. (2016). Calculation of gamma-ray attenuation parameters for locally developed shielding material: Polyboron. Journal of Radiation Research and Applied Sciences, 9, 26-34.

Gerward, L., Guilbert, N., Jensen, K.B., Levring, H., (2004). WinXCom—a program for calculating X-ray attenuation coefficients. Radiation Physics and Chemistry, 71, 653–654.

Downloads

Published

2023-12-08

How to Cite

Sreebunpeng, K., Chewpraditkul, W., Kimpraphan, N., Chimalawong, P., & Wongwigkarn, K. (2023). Attenuation Coefficient and Mean Free Path of 60, 290, 404, 566 and 662 keV -rays for PbWO4 and CdWO4 Scintillation Crystals. Science Journal, Chandrakasem Rajabhat University, 33(2), 74–82. retrieved from https://ph03.tci-thaijo.org/index.php/scicru/article/view/541

Issue

Section

Research Articles